IFM TR8430
Request a Quote
Quick facts
| Availability | Confirmed on request |
| Warranty | 12-month (new) · 3-month (used/refurbished) |
Shipping & availability
When can TR8430 be shipped?
TR8430 is sourced on request — request a quote and we'll confirm the exact availability and shipping timeline for this specific unit.
Can I pick up TR8430 in person?
Local pickup from our Groningen, Netherlands warehouse may be possible — contact us before ordering to check availability and arrange a collection time.
Is express delivery available?
Yes — mention it when you request a quote for TR8430 and we'll include express shipping options in our reply.
My factory is down and I need this urgently — can you help?
For a critical breakdown, contact us directly at +31 (0)50-7110348 or sales@partnerplc.com — we can look into same-day courier or dedicated taxi/hotshot delivery for TR8430, subject to availability and destination.
Wide temperature evaluation range With four programmable switching outputs Intuitive, user-friendly handling Local display of the current temperature values Automatic recognition of the connected sensor type
Specifications
Product characteristics
| Number of inputs and outputs | Number of digital outputs: 4 |
| Measuring range | -40...150 °C; -40...302 °F |
| -40...150 °C | -40...302 °F |
| Process connection | threaded connection G 1/2 external thread |
Application
| Special feature | Gold-plated contacts |
| Application | for Pt100 and Pt1000 measuring elements |
Electrical data
| Operating voltage[V] [V] | 18...28 DC; (to SELV/PELV) |
| Current consumption[mA] [mA] | < 90; (when display is active) |
| Protection class | III |
| Reverse polarity protection | yes |
| Power-on delay time[s] [s] | 1.5 |
| Integrated watchdog | yes |
Inputs / outputs
| Number of inputs and outputs | Number of digital outputs: 4 |
Outputs
| Total number of outputs | 4 |
| Output signal | switching signal |
| Electrical design | PNP |
| Number of digital outputs | 4 |
| Output function | normally open / normally closed; (parameterisable) |
| Max. voltage drop switching output DC[V] [V] | 2 |
| Permanent current rating of switching output DC[mA] [mA] | 500; (max. value for each single output) |
| Short-circuit protection | yes |
| Type of short-circuit protection | pulsed |
| Overload protection | yes |
Measuring/setting range
| Measuring range | -40...150 °C; -40...302 °F |
| -40...150 °C | -40...302 °F |
| Set point SP | -39.8...150 °C; -39.5...302 °F |
| -39.8...150 °C | -39.5...302 °F |
| Reset point rP | -40...149.8 °C; -40...301.5 °F |
| -40...149.8 °C | -40...301.5 °F |
| In steps of | 0.2 °C; 0.5 °F |
| 0.2 °C | 0.5 °F |
| Resolution of switching output[K] [K] | 0.2 |
| Resolution of display[K] [K] | 0.2 |
Accuracy / deviations
| Switch point accuracy[K] [K] | ± 0,2 |
| Display accuracy[K] [K] | ± (0,2 + ½ Digit) |
| Temperature drift per 10 K[K] [K] | 0.1 |
| Temperature coefficient[% of the span / 10 K] [% of the span / 10 K] | in case of deviation from the reference condition 25 ± 5 °C |
Response times
| Response time[ms] [ms] | 200 |
Software / programming
| Adjustment of the switch point | programming button |
| Parameter setting options | hysteresis / window; normally open / normally closed; min/max memory reset; Display unit; zero-point calibration |
Operating conditions
| Ambient temperature[°C] [°C] | -25...70 |
| Storage temperature[°C] [°C] | -40...85 |
| Protection | IP 67 |
Tests / approvals
| EMC | DIN EN 61000-4-2 ESD; 4 kV CD / 8 kV AD; EN 61000-4-3 HF radiated; 10 V/m; DIN EN 61000-4-4 Burst; 2 kV; EN 61000-4-6 HF conducted; 10 V |
| DIN EN 61000-4-2 ESD | 4 kV CD / 8 kV AD |
| EN 61000-4-3 HF radiated | 10 V/m |
| DIN EN 61000-4-4 Burst | 2 kV |
| EN 61000-4-6 HF conducted | 10 V |
| Shock resistance | DIN IEC 68-2-27; 50 g (11 ms) |
| DIN IEC 68-2-27 | 50 g (11 ms) |
| Vibration resistance | DIN EN 60068-2-6; 20 g (10...2000 Hz) |
| DIN EN 60068-2-6 | 20 g (10...2000 Hz) |
| MTTF[years] [years] | 175 |
Mechanical data
| Weight[g] [g] | 265.5 |
| Housing | cylindrical |
| Dimensions[mm] [mm] | Ø 34 / L = 93 |
| Materials | stainless steel (304/1.4301); stainless steel (303/1.4305); EPDM/X; PC; PBT; FKM |
| Process connection | threaded connection G 1/2 external thread |
Displays / operating elements
| Display | Display unit; 2 x LED, green; switching status; 4 x LED, yellow; function display; alphanumeric display, 4-digit; measured values; alphanumeric display, 4-digit |
| Display unit | 2 x LED, green |
| switching status | 4 x LED, yellow |
| function display | alphanumeric display, 4-digit |
| measured values | alphanumeric display, 4-digit |
Remarks
| Remarks | n.c. = not used; referring to UL: for use on a low voltage circuit with overcurrent protection; in accordance with UL873 tab. 28.1 or Imax = 100/Ub (Ub = voltage of the circuit); total current rating of the unit: max. 1 A; Two outputs can be switched in parallel to achieve a total current rating of 1 A.; Switch-on points and switch-off points and the programming of the outputs must correspond exactly. |
| Pack quantity | 1 pcs. |
Electrical connection
| Connector | 1 x M12; coding: A; Contacts: gold-plated |